Send Email
 

Bilirubin Phototherapy Analog Sensor PMA 1123 

 

The PMA1123 Bilirubin Phototherapy Analog Sensor is an accurate, stable sensor designed to measure the output of phototherapy lamps used in neonatal wards to treat infant hyperbilirubenenima (jaundice). Blue light in the spectral range of 425-475 nm chemically alters bilirubin below the surface of the skin into products the infant can eliminate in urine.

> Return to Sensor Selection Guide

U.S.Pat.5,946,641

U.S.Pat.5,790,432

FEATURES

  • High sensitivity
  • Wide dynamic range
  • Excellent long term stability
  • Cosine corrected
  • NIST traceable calibration
  • Selectable Units

APPLICATIONS

  • Phototherapy Lamp monitoring
  • Determining length of exposure
  • Clinical studies

> Weatherproof UV Detector available

> Underwater UV Detector available

> Wiring Diagram

 


The standard PMA1123 analog sensor has a narrow spectral response from 425-475 nm, closely matching the action spectrum for bilirubin breakdown. The measurement will be accurate for the output of any lamp, in particular, blue fluorescent, daylight fluorescent and quartz halogen lamps commonly used in bilirubin light units. The PMA1123 detector is calibrated to read in µW/cm2.

Specifications:

  • Spectral response: 425-475 nm

  • Range: 2,000 [μW/cm2] (Standard) ; 20,000 [μW/cm2] (PMA1123C)

  • Display resolution: 0.01 [μw/cm2]; 0.1[μW/m2] (PMA1123C)

  • Operating environment: 32 to 120 °F (0 to +50 °C) no precipitation

  • Cable:  5ft (1.5m)

  • Diameter: 1.6" (40.6 mm)

  • Height: 1.8" (45.8 mm)

  • Weight: 7.1 oz. (200 grams)

> Return to Sensor Selection Guide

© 2011 Solar Light. All rights reserved.