The PMA4100 Series
Photostability detectors are process control detectors that have the
capability of being calibrated in the field by the end user.
The PMA 4100-420
Detector System has two detector types, the PMA 4110-420 UVA Detector
and the PMA 4130-420 Photopic Detector. The detectors detect light from
their respective spectral bands and send a current proportioned signal
on a 4 – 20 mA current loop. The PMA 4110-420 Detector is sensitive to
the 320 – 400 nm spectral band and has a maximum (20 mA) sensitivity of
1.000 W/cm². The PMA 4130-420 Detector is sensitive to the CIE Weighted
Photopic spectral range, 480 - 620 nm, and has a maximum (20 mA)
sensitivity of 5000 Lux.
This detector system
includes reference detectors, transfer calibrated from NIST traceable
standards. One Reference detector is available to calibrate the Process
detector for the Photopic range and one Reference detector is available
to calibrate the Process detector for the UVA. The software system can
perform an in situ transfer calibration when connected via the
USB.
> Return to Detector Selection Guide |
.jpg)
|
FEATURES
- High sensitivity
- Dynamic range 2*105
- Excellent long term stability
- Cosine corrected
- NIST traceable calibration
- Radiometric units and max. exposure time
- Field Calibration
APPLICATIONS
|
|
SL3310 Calibration
software
The unique feature of infield calibration is capable
through use of the SL3301 software that can be installed on a laptop
computer
to meet performance standards.
|
Photopic Detector Specifications:
- Spectral response: Follows CIE photopic spectral luminous efficiency curve (400-700nm) Figure 1
- Angular response: 5% for angles <60°
- Range:
PMA2130 - 150,000 Lux, 220 W/m2, 14,000 ft-cd
PMA2130L - 1,500 Lux, 2.2 W/m2, 140 ft-cd
PMA2130H - 1,500,000 Lux, 2200 W/m2, 140,000 ft-cd
- Display resolution:
PMA2130 - 1 Lux, 0.001 W/m2, 0.1 ft-cd
PMA2130L - 0.01 Lux, 0.00001 W/m2, 0.001ft-cd
PMA2130H - 10 Lux, 0.01 W/m2, 1 ft-cd
- Operating environment: 32 to 120 °F (0 to +50 °C) no precipitation
- Cable: 5ft (1.5m)
- Diameter: 1.6" (40.6 mm)
- Height: 1.8" (45.8 mm)
- Weight: 7.1 oz. (200 grams)
UVA Detector Specifications:
- Spectral response:
320-400nm, figure 1
- Angular response:
5% for angles <60°
- Range: 200 [mW/cm2] or
2000 [W/m2]
- Display resolution: 0.001 [mw/cm2],
0.01[W/m2]
- Operating environment:
32 to 120 °F (0 to +50 °C) no precipitation
- Temperature coefficient: <0.1%/°C
- Cable: 5ft (1.5m)
- Diameter: 1.6" (40.6 mm)
- Height: 1.8" (45.8 mm)
- Weight: 7.1 oz. (200
grams)